In situ measurement of tissue impedance using an inductive coupling interface circuit

  • Hung Wei Chiu
  • , Jia Min Chuang
  • , Chien Chi Lu
  • , Wei Tso Lin
  • , Chii Wann Lin
  • , Mu Lien Lin

研究成果: 期刊貢獻文章同行評審

6 引文 斯高帕斯(Scopus)

摘要

In this work, a method of an inductive coupling impedance measurement (ICIM) is proposed for measuring the nerve impedance of a dorsal root ganglion (DRG) under PRF stimulation. ICIM provides a contactless interface for measuring the reflected impedance by an impedance analyzer with a low excitation voltage of 7 mV. The paper develops a calibration procedure involving a 50-Ω reference resistor to calibrate the reflected resistance for measuring resistance of the nerve in the test. A de-embedding technique to build the equivalent transformer circuit model for the ICIM circuit is also presented. A batteryless PRF stimulator with ICIM circuit demonstrated good accuracy for the acute measurement of DRG impedance both in situ and in vivo. Besides, an in vivo animal experiment was conducted to show that the effectiveness of pulsed radiofrequency (PRF) stimulation in relieving pain gradually declined as the impedance of the stimulated nerve increased. The experiment also revealed that the excitation voltage for measuring impedance below 25 mV can prevent the excitation of a nonlinear response of DRG.

原文English
文章編號6226810
頁(從 - 到)225-235
頁數11
期刊IEEE Transactions on Biomedical Circuits and Systems
7
發行號3
DOIs
出版狀態Published - 2013

指紋