跳至主導覽 跳至搜尋 跳過主要內容

Disposable redox polymer coated screen-printed carbon electrode for NADH sensing

研究成果: 期刊貢獻文章同行評審

1 引文 斯高帕斯(Scopus)

摘要

Electrochemical behavior of electro-generated poly-pyronin B (PyB) film was reported at a screen-printed carbon electrode (SPCE). The poly(PyB) modified SPCE showed excellent redox activity in neutral and alkaline media. Surface topography of poly(PyB) film modified electrode was analyzed by using atomic force microscopy (AFM) and scanning electron microscopy (SEM). Poly(PyB) film coated electrode was characterized by electrochemical impedance spectroscopy (EIS), UV-visible absorption spectroscopy (UV-vis) and cyclic voltammetry (CV). The poly(PyB) modified electrode showed electrocatalytic response to the reduced form of nicotinamide adenine dinucleotide (NADH) in physiological condition, and was used for the detection of NADH with high selectivity. The anodic peak current was linearly related to concentrations of NADH over the range from 1 × 10-5 M to 5.2×10-4 M, and the detection limit was 5×10-7 M (S/N = 3).

原文English
頁(從 - 到)172-179
頁數8
期刊Micro and Nanosystems
4
發行號3
DOIs
出版狀態Published - 2012

指紋

深入研究「Disposable redox polymer coated screen-printed carbon electrode for NADH sensing」主題。共同形成了獨特的指紋。

引用此