摘要
The direct electrochemistry of glucose oxidase (GOx) was successfully realized on electrochemically reduced graphene oxide and silver nanoparticles (RGO/Ag) nanocomposite modified electrode. The fabricated nanocomposite was characterized by field emission scanning electron microscope and energy dispersive spectroscopy. The GOx immobilized nanocomposite modified electrode showed a pair of well-defined redox peaks with a formal potential (E°) of -0.422V, indicating that the bioactivity of GOx was retained. The heterogeneous electron transfer rate constant (Ks) of GOx at the nanocomposite was calculated to be 5.27s-1, revealing a fast direct electron transfer of GOx. The GOx immobilized RGO/Ag nanocomposite electrode exhibited a good electrocatalytic activity toward glucose over a linear concentration range from 0.5 to 12.5mM with a detection limit of 0.16mM. Besides, the fabricated biosensor showed an acceptable sensitivity and selectivity for glucose.
| 原文 | English |
|---|---|
| 頁(從 - 到) | 164-169 |
| 頁數 | 6 |
| 期刊 | Colloids and Surfaces B: Biointerfaces |
| 卷 | 114 |
| DOIs | |
| 出版狀態 | Published - 1 2月 2014 |
文獻附註
Funding Information:This project was supported by the National Science Council and the Ministry of Education of Taiwan (Republic of China) .